✨ NewHighSanqiang endoscope cleaning workstation
Multi-station endoscope cleaning line in a straight, L- or U-shaped layout with final drying.
Cleaning determines the success of the entire process — protein residue on an instrument makes sterilization impossible regardless of the method used. Decontamination therefore begins with a washer-disinfector providing thermal disinfection and a documented A0 value, supplemented for complex geometries by an ultrasonic washer that removes soil from holes and channels a water jet cannot reach. Flexible endoscopes require a separate washer compliant with endoscope reprocessing standards.
A0 expresses thermal disinfection efficacy as exposure time normalised to 80°C. A0 ≥ 600 is assumed for surgical instruments, and A0 ≥ 3000 for devices contacting broken skin or mucous membranes.
No. Ultrasound removes soil mechanically from hard-to-reach places but provides no thermal disinfection. It is a stage preceding the main wash, not a substitute for it.
Yes. The pass-through version with doors on both sides separates the dirty and clean zones, which is standard in central sterile supply departments and eliminates the risk of re-contaminating the load.
4 questions — and we will show matching items from our catalogue
✨ NewHighMulti-station endoscope cleaning line in a straight, L- or U-shaped layout with final drying.
✨ NewExpertAutomatic flexible endoscope washer-disinfector per WS 507, GB/T 3526 and GB 30689.
✨ NewBasicUltrasonic washer for instruments and optics — cavitation removes soil from lumens and blind holes.
✨ NewHigh350 l washer-disinfector with multiple washing, disinfection and drying modes.
✨ NewExpertFast washer-disinfector, 470 l, 6 programmes, 35-minute enzymatic cycle, pass-through doors.
✨ NewExpertFast washer-disinfector, 536 l, 6 programmes, 35-minute enzymatic cycle, pass-through doors.
A0 expresses thermal disinfection efficacy as exposure time normalised to 80°C. A0 ≥ 600 is assumed for surgical instruments, and A0 ≥ 3000 for devices contacting broken skin or mucous membranes.
No. Ultrasound removes soil mechanically from hard-to-reach places but provides no thermal disinfection. It is a stage preceding the main wash, not a substitute for it.
Yes. The pass-through version with doors on both sides separates the dirty and clean zones, which is standard in central sterile supply departments and eliminates the risk of re-contaminating the load.